Description
The ideal portable pressure calibrator, the 729 allows you to simply type in a target pressure, and the calibrator will automatically pump to the desired set-point. Then, internal fine adjustment conÂtrol automatically stabilizes the pressure at the requested value.
The Fluke 729 can also automatically test multiple pressure test points and automatically document the results. Calibration is as easy as typing in the starting and ending pressure and the number of test points and tolerance level. The 729 does the rest.
Built-in HART communication capabilities enable HART transmitter mA adjustments, light HART configuration and the ability to adjust to applied 0% and 100% values.
Upload and manage documented calibration results with DPCTrack2â„¢ Calibration Management Software, making it easy to manage your instrumentation, create scheduled tests and reports, and manage calibration data.
With three ranges to choose from, 30 psi (2 bar, 200 kPa), 150 psi (10 bar, 1 MPa) and 300 psi (20 bar, 2 MPa), Fluke 729 Automatic Pressure Calibrators are designed to perform when and where you need them.
Key features
- Automatic pressure generation and regulation to 300 psi
- Easily document the process using onboard test templates
- Automatic internal fine-pressure adjustment
- Measure, source and simulate 4 to 20 mA signals
- 24V loop power for powering transmitters for tests
- Bright dual/triple-channel color graphical display
- HART communication for testing HART smart transmitters
- Built-in leak testing for verifying test setups for leakage
- Rugged portable design with standard 3-year warranty
- Fluke Connect® App compatibility
- Measure temperature with optional 720RTD probe
- Measure Vdc for testing 1 V dc to 5 V dc output transmitters
- Compatible with 700 Series and 750 Series pressure modules
- Upload documented test results to DPCTrack2â„¢ Calibration Management Software
- Innovative semi-rigid carrying case designed for field work
Automatic pressure generation and control to 300 psi (20 bar, 2 MPa). Enter a pressure and the 729 automatically pumps to the desired pressure. Fill in a test template and the 729 will automatiÂcally pump to and document a multiple-point pressure calibration test.
Easy calibration documentation using defined templates for transmitters and switches. Type in the starting and ending test pressures and number of test points. The 729 does the rest by documenting the applied pressure, measured mA and % error for each test point. The bright color graphical display will flag out-of-tolerance test results in red.
Automatic internal fine pressure adjustment helps compensate for minor leaks in hoses and test setups. No need to manually turn a vernier to compensate for leaks helping to eliminate the hassle of manual pumping and manual fine adjustment of test setups.
HART communication enables mA output trim, trim to applied values and pressure zero trimming of HART pressure transmitÂters. You can also do light configuration tasks such as changing a transmitter tag, measurement units and ranging. Other supported HART commands include setting fixed mA outputs for troubleÂshooting, read device configuration and variables and read device diagnostics.
Measure mA signals on transmitter outputs, source and simulate mA signals for testing I/Ps and other mA loop devices. Includes a 24V loop power supply for testing and powering transÂmitters in standalone tests disconnected from the control system.
Rugged portable design and 3-year warranty set the 729 apart from most other pressure calibrators. The 729 is tested and warÂranted to withstand a 1-meter drop test so it is ready for field instrumentation calibration work.
Fluke Connect® compatibility enables a user to perform remote monitoring using the Fluke Connect mobile app and also manage and store pressure measurements and logging events. Share meaÂsurements using ShareLiveâ„¢ video calls and emails.
Measure temperature with optional 720RTD probe for performing custody transfer calibrations.
Measure pressure with legacy 700 Series and the 750 Series pressure modules, if different pressure measurement ranges or resolution is desired.
Multiple language support: select language from the instrument setup utility. Hart communication commands are limited to English per the HART device descriptors and protocol.
Semi-rigid carrying case designed for field work, with room in the case to store test hoses, fittings, test leads and small hand tools needed for field pressure calibrations.
Specifications
Pressure specification* |
One-year specification 0.02% of full scale |
Control specification 0.005% full scale minimum |
*See manual for detailed specifications |
Model |
729 30G |
Psi range, resolution |
-12.0000 to 30.0000 psi |
Bar range, resolution |
-0.82737 to 2.06842 bar |
kPa range, resolution |
-82.737 to 206.843 kPa |
Comment |
No wireless communication |
|
729 150G |
Psi range, resolution |
-12.000 to 150.000 psi |
Bar range, resolution |
-0.8273 to 10.3421 bar |
kPa range, resolution |
-82.73 to 1034.21 kPa |
Comment |
No wireless communication |
|
729 300G |
Psi range, resolution |
-12.000 to 300.000 psi |
Bar range, resolution |
-0.8273 to 20.6843 bar |
kPa range, resolution |
-82.73 to 2068.43 kPa |
Comment |
No wireless communication |
|
729 30G FC |
Psi range, resolution |
-12.0000 to 30.0000 psi |
Bar range, resolution |
-0.82737 to 2.06842 bar |
kPa range, resolution |
-82.737 to 206.843 kPa |
Comment |
Wireless communication for Fluke Connect |
|
729 150G FC |
Psi range, resolution |
-12.000 to 150.000 psi |
Bar range, resolution |
-0.8273 to 10.3421 bar |
kPa range, resolution |
-82.73 to 1034.21 kPa |
Comment |
Wireless communication for Fluke Connect |
|
729 300G FC |
Psi range, resolution |
-12.000 to 300.000 psi |
Bar range, resolution |
-0.8273 to 20.6843 bar |
kPa range, resolution |
-82.73 to 2068.43 kPa |
Comment |
Wireless communication for Fluke Connect |
|
729CN 200K |
Psi range, resolution |
-12.0000 to 30.0000 psi |
Bar range, resolution |
-0.82737 to 2.06842 bar |
kPa range, resolution |
-82.737 to 206.843 kPa |
Comment |
For China, no wireless communication |
|
729CN 1M |
Psi range, resolution |
-12.000 to 150.000 psi |
Bar range, resolution |
-0.8273 to 10.3421 bar |
kPa range, resolution |
-82.73 to 1034.21 kPa |
Comment |
For China, no wireless communication |
|
729CN 2M |
Psi range, resolution |
-12.000 to 300.000 psi |
Bar range, resolution |
-0.8273 to 20.6843 bar |
kPa range, resolution |
-82.73 to 2068.43 kPa |
Comment |
For China, no wireless communication |
|
729CN 200K FC |
Psi range, resolution |
-12.0000 to 30.0000 psi |
Bar range, resolution |
-0.82737 to 2.06842 bar |
kPa range, resolution |
-82.737 to 206.843 kPa |
Comment |
For China, wireless communication for Fluke Connect |
|
729CN 1M FC |
Psi range, resolution |
-12.000 to 150.000 psi |
Bar range, resolution |
-0.8273 to 10.3421 bar |
kPa range, resolution |
-82.73 to 1034.21 kPa |
Comment |
For China, wireless communication for Fluke Connect |
|
729CN 2M FC |
Psi range, resolution |
-12.000 to 300.000 psi |
Bar range, resolution |
-0.8273 to 20.6843 bar |
kPa range, resolution |
-82.73 to 2068.43 kPa |
Comment |
For China, wireless communication for Fluke Connect |
|
729JP 200K |
kPa range, resolution |
-82.737 to 206.843 kPa |
Comment |
For Japan, no wireless communication |
|
729JP 1M |
kPa range, resolution |
-82.73 to 1034.21 kPa |
Comment |
For Japan, 1 MPa range, no wireless communication |
|
729JP 2M |
kPa range, resolution |
-82.73 to 2068.43 kPa |
Comment |
For Japan, 2 MPa range, no wireless communication |
|
729JP 200K FC |
kPa range, resolution |
-82.737 to 206.843 kPa |
Comment |
For Japan, wireless communication for Fluke Connect |
|
729JP 1M FC |
kPa range, resolution |
-82.73 to 1034.21 kPa |
Comment |
For Japan, wireless communication for Fluke Connect |
|
729JP 2M FC |
kPa range, resolution |
-82.73 to 2068.43 kPa |
Comment |
For Japan, wireless communication for Fluke Connect |
|
Electrical specification |
All specifications are valid to 110% of range, except 24 mA source and simulate which are valid to 100% of range. |
Ranges |
mA measure, source and simulate |
0 mA to 24 mA |
Volts dc measure |
0 V dc to 30 V dc |
Resolution |
mA dc source, simulate and measure |
1 μA |
Voltage dc measurement |
1 mV |
Accuracy |
0.01% ±2 LSD all ranges (at 23 °C ±5 °C) |
Stability |
20 ppm of full scale /°C from -10 °C to +18 °C and 28 °C to +50 °C |
mA simulate external voltage requirement |
12 V dc to 30 V dc |
Loop compliance voltage |
24 V dc @ 20 mA |
mA simulate external voltage requirement |
12 V dc to 30 V dc |
Temperature measurement only/100 Ω Pt(385) RTD |
-50 °C to 150 °C (-58 °F to 302 °F) |
Temperature resolution |
0.01 °C (0.01 °F) |
Temperature accuracy |
±0.1 °C (0.2 °F) ±0.25 °C (±0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory) |
Drive capability |
1200 Ω without HART resistor, 950 Ω with internal HART resistor |
Mechanical specification |
Size (HxWxL) |
7 x 27.94 x 17.27 cm (2.75 x 11 x 6.8 in) |
Weight |
6.5 lbs., 2.95 kg |
Ingress protection |
IP54 |
Environmental specification |
Operating temperature |
-10 °C to 50 °C for measurement, 0 °C to 50 °C for pressure control |
Operating temperature with battery |
-10 °C to 40 °C |
Battery will only charge to 40 °C |
Storage temperature |
-40 °C to 60 °C |
Operating altitude |
3000 m |
Storage altitude |
13000 m |
Wireless (729 FC only) |
Radio frequency certification |
FCC ID: T68-FBLE IC:6627A-FBLE |
Wireless radio frequency range |
2412 MHz to 2462 MHz |